Self-tests, Auto tests, Table 10-1 – Physio-Control LIFEPAK 15 User Manual

Page 210: General maintenance and testing

Advertising
background image

10-4

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Maintenance and Testing

Self-Tests

Each time you turn on the LIFEPAK 15 monitor/defibrillator, it performs internal self-tests to check

that internal electrical components and circuitry work properly. The defibrillator stores the results of

all user-initiated self-tests in a test log.

When the defibrillator is on and a problem is detected that requires immediate service, such as a

malfunctioning charging circuit, the Service LED is illuminated.

For more information, see Table 10-2 on page 10-18.

Auto Tests

The defibrillator performs an automatic self-test daily at 03:00 (3:00 A.M.), if not in use. During

the automatic self-test, the defibrillator turns itself on (

ON

LED illuminates) briefly and completes

the following tasks:

• Performs a self-test

• Stores the self-test results in the test log

• Prints the self-test results

Table 10-1 Recommended Maintenance Schedule for Clinical Personnel

OPERATION

DAILY

AFTER

USE

AS

REQUIRED

6

MONTHS

12

MONTHS

Complete Operator’s Checklist. Includes
QUIK-COMBO therapy cable check and
Standard Paddles Monitoring and User Test

X

Inspect defibrillator

X

X

Check that all necessary supplies and
accessories are present (for example, fully
charged batteries, gel, electrodes, ECG paper,
etc.)

X

X

X

Function Checks:

Patient ECG Cable Check

X

Standard Paddles Synchronized
Cardioversion Check

X

Therapy Cable Monitoring and
Synchronized Cardioversion Check

X

Therapy Cable Pacing Check

X

Clean defibrillator

X

X

Preventive Maintenance and Testing

X