Synchronized cardioversion procedure – Physio-Control LIFEPAK 15 User Manual

Page 144

Advertising
background image

5-30

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Synchronized Cardioversion Procedure

DISARMING

message appears

(shock)

button not pressed

within 60 seconds after charge
complete

• Recharge the defibrillator, if

desired.

Energy selected after charge
complete

• Recharge the defibrillator.

SPEED DIAL

pressed

• Recharge the defibrillator.

PACER

pressed

• Recharge, if necessary, or

no action, if pacing desired.

Therapy electrodes or cable
disconnected

• Reconnect electrode or

cable.

Energy did not escalate
automatically per energy
protocol

ENERGY SELECT

pressed and

disabled automatic protocol

• Continue to select energy

manually to treat patient.
For more information about
energy protocol, see
LIFEPAK 15 Monitor/
Defibrillator Setup Options

provided with your device.

SYNC

mode will not activate

PACER

is on. Pacing and Sync

are separate functions and are
not allowed at the same time.

• Discontinue pacing, if

appropriate for the patient,
and press SYNC.

ECG electrodes not attached to
patient and standard paddles
connected to defibrillator

• Connect ECG electrodes to

patient.

Patient did not “jump” (no
muscle response) during
defibrillator discharge

Patient muscle response is
variable and depends on patient
condition. Lack of visible
response to defibrillation does
not necessarily mean the
discharge did not occur.

• No action needed.

The Test Load is connected to
therapy cable

• Remove the Test Load and

connect therapy electrodes
to cable.

Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION