PADDLE ACCESSORY OPTIONS

6

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

6-9

Figure 6-8 Anterior-Lateral Paddle Position

Anterior-Posterior Placement

Place the

STERNUM

paddle anteriorly over the left precordium and

the

APEX

paddle posteriorly behind the heart in the infrascapular area (see Figure 6-9).

Figure 6-9 Anterior-Posterior Paddle Position

Cleaning Pediatric Paddles

Individually protect paddles before and after cleaning to prevent damage to paddle surfaces.

After each use:

1. Wipe or rinse paddle electrodes, cable connector, paddle handles, and cables with mild soap

and water or disinfectant using a damp sponge, towel, or brush. Do not immerse or soak.

2. Dry thoroughly.

3. Examine paddle surfaces, connector, handles, and cables for damage or signs of wear.

• Cables that show signs of wear such as loose cable connections, exposed wires, or cable

connector corrosion should be removed from use immediately.

• Paddles that have rough or pitted electrodes should be removed from use immediately.

Sternum

Apex

Sternum

Apex