General troubleshooting tips, General troubleshooting – Physio-Control LIFEPAK 15 User Manual

Page 224

Advertising
background image

10-18

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Troubleshooting Tips

General Troubleshooting Tips

If a problem is detected with the LIFEPAK 15 monitor/defibrillator during operation or testing, refer

to the troubleshooting tips in Table 10-2. If the problem cannot be corrected, remove the

LIFEPAK 15 monitor/defibrillator from active use and contact a qualified service technician for

service and repair.

Table 10-2 General Troubleshooting Tips

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

No power when monitor/
defibrillator is turned ON

Low battery voltage

• Replace with fully charged, properly

maintained battery.

Battery connector pin loose,
covered with foreign
substance, or damaged

• Remove battery and inspect pins.

Clean if foreign substance present.
Contact a qualified service
technician to replace if bent,
cracked, or loose.

Power adapter not properly
connected to auxiliary power
source

• Check that power adapter is properly

connected to auxiliary power.

Power adapter not properly
connected to monitor/
defibrillator

• Check that power adapter is properly

connected to monitor/defibrillator.

Defective power adapter or
cables

• Replace with working power adapter

and cables.

Defective battery

• Remove battery from service and

replace with working battery.

ON

Led illuminated, but

screen is blank and device
does not operate

Device boot up has failed

• Press and hold ON until LED turns

off (~5 seconds). Then press ON to
turn device back on.

• If device does not turn off, remove

both batteries and disconnect device
from power adapter, if applicable.
Then reinsert batteries, reconnect
power adapter, and press ON to turn
device back on.