2 analog input resistance calibration – Advantech Data Acquisition Modules ADAM 4000 User Manual

Page 317

Advertising
background image

Chapter

8

Chapter 8 Calibration 8-5

8.2 Analog Input Resistance Calibration

Model: ADAM-4013

1. Apply power to the module and let it warm up for about 30 minutes.

2. Assure that the module is correctly installed and is properly

configured for the input range that you want to calibrate. You can do
this by using the ADAM utility software. (Please refer to Appendix
D, Utility Software.)

3. Apply the reference Span resistance to the screw terminals using a

4-wire connection. Please refer to Figure 8-5 for the correct wiring
diagram and Table 8-2 for the right Span calibration resistance).

Figure 8-5 Applying calibration resistance


Use a precision resistance decade box or discrete resistors with values:
10 W, 15 W, 60 W, 140 W, 200 W and 440 W.

4. Apply the reference Offset resistance to the terminals of the module.

(Please refer to Figure 8-5 for the correct wiring diagram and Table
8-2 for the right Offset calibration resistance.)

5. Issue an Offset Calibration command to the module. This can be

done by using the Calibrate option in the ADAM utility software
(Please see Appendix D, Utility Software)

6. Issue a Span Calibration command to the module. This can be done

through the Calibrate option in the ADAM utility software (Please
see Appendix D, Utility Software)

Advertising