Calibration


8-6 ADAM 4000 Series User’s Manual

Table 8-2 Calibration Resistance

Module

Input
Range
Code
(Hex)

Input Range

Span
Calibration
Resistance

Offset
Calibration
Resistance

4013 20

Pt, -100~100 ℃

a = 0.00385

140

Ω 60

Ω

21

Pt, 0~100 ℃

a = 0.00385

140

Ω 60

Ω

22

Pt, 0~200 ℃

a = 0.00385

200

Ω 60

Ω

23

Pt, 0~600 ℃

a = 0.00385

440

Ω 60

Ω

24

Pt, -100~100 ℃

a = 0.003916

140

Ω 60

Ω

25

Pt, 0~100 ℃

a = 0.003916

140

Ω 60

Ω

26

Pt, 0~200 ℃

a = 0.003916

200

Ω 60

Ω

27

Pt, 0~200 ℃

a = 0.003916

440

Ω 60

Ω

28

Ni, -80~100 ℃

200

Ω 60

Ω

29

Ni, 0~100 ℃

200

Ω 60

Ω