Luminex xPONENT 4.0 Software Manual

92

Calibration expiration days - Type the desired number of days a system calibration
remains active before expiring. The default number of days is seven.

Analysis display decimal places - Type the number of digits the system displays.
The default number of digits is three.

Minimum bead count for obtaining results - Type any whole number from 0 to
1000000. This is the minimum bead count that must be observed by the instrument for

a particular analyte before this data is used in statistical calculations and is displayed

on the graphs and results table. When set to a number greater than 0, the analyzer

does not display data for bead sets not generating events equal to or less than this

value. Type 0 to display all events.

Allow batches to be run or saved without lot number, expiration or manufacturer
-
Allows saving lots without normal required information.

Run Routine after batch completes - Select this option to specify a particular routine
to run after a batch is finished. If you select this option, then select a routine from the

Routine to run after batch completes list.

Batch Thresholds. Displays the following options:

Low bead count detection - Selecting either Error or Warning stops the batch or
logs a warning if the number of successive wells in the Well Count box are run without

reaching the number of beads in the Bead Count box. Click in either box to type a

number other than the default.

Liquid Detection - Selecting either Error or Warning stops the batch or logs a
warning if no liquid is detected in the well.

Cancel. Cancels changes.

Save. Saves changes.