Perform system calibration, Manage maintenance routines, Create new routine – Luminex xPONENT 4.0 SP1 User Manual

Page 139: Edit routine

Advertising
background image

127

Using the Software

WARNING:

Lack of proper maintenance can lead to inaccurate test results
and failure of system components. Please follow all scheduled
maintenance routines to ensure proper functioning of equipment.

Perform System Calibration

Perform system calibration weekly. In addition, recalibrate the system if the delta cal
temperature exceeds +/-5ºC. For further information, see “System Initialization” on
page 108
.

Manage Maintenance Routines

You manage maintenance routines at the Cmds & Routines tab of the Maintenance
page. Open the Maintenance page, then click the Cmds & Routines tab.

Create New Routine

Create maintenance routines to facilitate your startup procedure, shutdown procedure,
troubleshooting, or calibration. Ensure that the maintenance routine you create is
consistent with the Luminex analyzer operations and maintenance requirements. See the
Luminex System Manual for more information.

1. Open the Maintenance page, then open the Cmds & Routines tab.
2. In the Routine Name list, click None.
3. In the Plate Name list, select the plate that you will use for the new routine.
4. In the Commands section, click one or more commands to add to the routine. These

commands display in the plate image and in the Command Sequence list.

5. To change a location (well or reservoir) for a command, click the command in the

Command Sequence list, then click the new location in the plate image.

NOTE:

If you attempt to place two or more incompatible commands in the
same well, a message will display instructing you to change the
location of the command. Some commands can be run from the
same location, for example, multiple washes can be run from the
reservoir.

6. Click Save As to save the new routine. The Save Routine dialog box opens.
7. Type the routine name in the Routine Name box, then click OK.

Edit Routine

You can also edit routines.

1. Open the Maintenance page, then open the Cmds & Routines tab. Click the routine

that you want to edit from the Routine Name list.

2. Click a command in the Command Sequence list that you wish to edit. Click Clear to

delete the command, or click in a different well in the plate image to change the
location of that command.

NOTE:

When you modify a routine, the routine name automatically
defaults to None in the Routine Name list.

3. Add, delete, or change commands as necessary, then click Save As.
4. Type a different name for the routine to create a new routine, or type the existing name

of the routine to keep the edited routine using the existing routine name.

Advertising