Luminex xPONENT 4.0 Software Manual

28

FIGURE 15.

Analysis Settings Dialog Box

If you selected Quantitative on the Settings tab, the default analysis formula is 5P
weighted. To change the default, select one of the following from the Method list:

No Analysis

Cubic Spline

Linear Fit

Logistic 4P

Logistic 5P

If you select Logistic 4P or Logistic 5P, select a weight type of either None or 1/Y

2

.

If you selected Qualitative on the Settings tab, the default analysis is Luminex
Qualitative
. Change the default value by selecting either Lum Qual or No Analysis.

Click Apply to All Analytes to apply your selection to all selected analytes. Click OK to
change the default analysis to the analysis you selected. Click Cancel to close the dialog
box without saving. For more information, see “Analysis Settings” on page 25.

Units. Type the desired units for the analytes in this box.

Count. Type the desired bead count for the analytes by clicking in the Count box. If each
selected bead set does not acquire this number of events, a warning is added to the log
that not enough bead events were acquired. If you select bead sets that are not present,
the analyzer continues to acquire, trying to reach the number of events per bead for bead
sets that are not in the sample. Therefore, choose only the bead sets present in your
sample.

Apply All. Applies the information in the Units and Count boxes to all analytes.

Advertising