Page x

P/N 12555-001 Rev 3

User Guide

Caution: United States federal law restricts this device to sale by or on the order of a licensed physician.

Caution: The AutoPulse is designed to be used only with ZOLL-approved accessories. The AutoPulse

will perform improperly if non-approved accessories are used.

Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other

batteries may cause permanent damage to the AutoPulse and will void the warranty.

Caution: Make sure that the LifeBand is not twisted before automatic compressions begin.

Caution: Do not submerge the AutoPulse Platform in liquid.

Caution: Use care while using sharp instruments around the LifeBand.

Caution: Do not block the vents of the AutoPulse.

Caution: Do not use the AutoPulse alone as a patient transportation aid.

Warning:

•

If a user advisory or fault cannot be cleared or a system error occurs during active operation,
immediately revert to manual CPR.

•

Do not strap across, or otherwise constrain, the LifeBand. Constraining the movement of the
LifeBand can damage or break the LifeBand.

•

Do not touch the patient while the AutoPulse is analyzing the patient’s size.

•

Failure to properly position a patient, both vertically and laterally with respect to the
AutoPulse Platform, may cause injury to the patient.

•

If you must move or realign the patient, you must press the Stop/Cancel button before
adjustment.

•

Do not place your hands or any other objects on or under the LifeBand while the AutoPulse
is analyzing the patient or during active operation.

•

Operating the AutoPulse on a patient for extended periods of time may result in minor skin
irritation to the patient.

•

Check the patient’s chest rise during ventilation during active operation.

•

Do not use the AutoPulse in the presence of an oxygen-rich (greater than 25% oxygen)
atmosphere, flammable anesthetics, or other flammable agents (such as gasoline). Using the
AutoPulse near the site of a gasoline spill may cause an explosion.

•

No modification of the AutoPulse Platform, the LifeBand, or the AutoPulse Power System
is allowed.