Multi-Chemistry Battery Charger with Charging Bay
2.
The Multi-Chemistry Battery Charger automatically detects the presence of both a AutoPulse Li-
Ion Battery and/or a AutoPulse NiMH Battery within 5 seconds.
3.
The Multi-Chemistry Battery Charger’s status will be indicated on the control panel (see Figure 4-2)
Figure 4-2
Multi-Chemistry Battery Charger Control Panel
4.
The Multi-Chemistry Battery Charger will automatically perform a Test-Cycle when needed. This
will automatically occur at a minimum of once every 30 days (For more information, refer to the
AutoPulse Power System User Guide).
Note:
Do not remove an AutoPulse Li-Ion or NiMH Battery from the Multi-Chemistry Battery
Charger until its charging completes or the Battery’s runtime will be reduced.
Note:
Do not remove an AutoPulse Li-Ion or NiMH Battery during a test-cycle or the Battery’s
runtime will be unknown. Removing a Battery during a Test-Cycle may cause the charger to
automatically enter a Test-Cycle mode the next time the Battery is inserted into it.