Page 4-2

P/N 12555-001 Rev 3

User Guide

Note:

Figure 4-1

Multi-Chemistry Battery Charger with Charging Bay

2.

The Multi-Chemistry Battery Charger automatically detects the presence of both a AutoPulse Li-
Ion Battery and/or a AutoPulse NiMH Battery within 5 seconds.

3.

The Multi-Chemistry Battery Charger’s status will be indicated on the control panel (see Figure 4-2)

Figure 4-2

Multi-Chemistry Battery Charger Control Panel

4.

The Multi-Chemistry Battery Charger will automatically perform a Test-Cycle when needed. This
will automatically occur at a minimum of once every 30 days (For more information, refer to the
AutoPulse Power System User Guide).

Note:

Do not remove an AutoPulse Li-Ion or NiMH Battery from the Multi-Chemistry Battery
Charger until its charging completes or the Battery’s runtime will be reduced.

Note:

Do not remove an AutoPulse Li-Ion or NiMH Battery during a test-cycle or the Battery’s
runtime will be unknown. Removing a Battery during a Test-Cycle may cause the charger to
automatically enter a Test-Cycle mode the next time the Battery is inserted into it.