1 user advisory (45) – ZOLL AutoPulse 100 User Manual

Page 72

Advertising
background image

Page 5-4

P/N 12555-001 Rev 3

User Guide

5.2.1

User Advisory (45)

Figure 5-5

User Advisory (45)

The AutoPulse driveshaft has a “home” position that is a point of reference for AutoPulse operation. If
the driveshaft is not at its home position when the AutoPulse is powered on, a User Advisory (45) will
occur. This User Advisory will persist until the driveshaft is returned to its home position.

To clear a User Advisory (45) pull up on the LifeBand until the chest bands are fully extended (thereby
moving the driveshaft back to its home position), and then RESTART.

If a User Advisory (45) can't be cleared; Power off the AutoPulse and contact ZOLL at +1.800.348.9011
or +1.978.421.9655

Warning: Removing the band clip when the driveshaft is not at its home position (see Figure 2-8 on

page 2-7) will result in a permanent User Advisory (45) that the user will not be able to
clear. To avoid this situation, the following guidelines should be adhered to:

1.

The LifeBand should be removed from the driveshaft ONLY from its home position.

2.

The LifeBand must be completely unwound with the seam visible (see Figure 2-10 on
page 2-8).

3.

If the LifeBand is cut, care should be taken to ensure that the bands are fully extended
before the cover plate is opened and the band clip is removed.

4.

Do not attempt to defeat the shaft lock mechanism, which keeps the driveshaft at its
home position when a LifeBand is not installed.

3 0 : 2