2 battery installation and removal – ZOLL AutoPulse 100 User Manual

Page 33

Advertising
background image

P/N 12555-001 Rev 3

Page 2-9

User Guide

2.2

Battery Installation and Removal

Remove a charged Battery from the AutoPulse Multi-Chemistry Battery Charger. Ensure the Battery is
fully charged before insertion into the AutoPulse Platform (refer to the “Performing a Battery Status
Check” section in the AutoPulse Power System User Guide for more information). Replace the Battery
remaining in the AutoPulse with one that is fully charged.

2.2.1

AutoPulse Li-Ion Battery Installation and Removal

Figure 2-11 AutoPulse Li-Ion Battery Installation and Removal

To install the AutoPulse Li-Ion Battery, first make sure the Battery Latch is rotated out of the way so that
the AutoPulse Li-Ion Battery can slide into the Battery compartment (see Figure 2-11). Then slide the
AutoPulse Li-Ion Battery into the Battery compartment in the AutoPulse Platform. The AutoPulse Li-
Ion Battery should snap into place and mount flush with the AutoPulse Platform. The Finger Latch for
the AutoPulse Li-Ion Battery should also be flush with the AutoPulse Platform such that the red strip on
the inside of the Finger Latch is not visible. Then rotate the Battery Latch into position as needed.

Finger Latch

Battery Latch Opened

Battery Latch Closed

Battery Latch

Locking Bar

Battery Front

Advertising