ZOLL AutoPulse 100 User Manual

Page 49

Advertising
background image

P/N 12555-001 Rev 3

Page 3-9

User Guide

2.

Press and release the Start/Continue button once. The AutoPulse automatically adjusts the bands to
the patient's chest (see Figure 3-9).

Figure 3-9

Analyzing Patient Size Display Panel Screen

3.

The AutoPulse will pause for 3 seconds to allow you to verify that the patient is properly aligned
and that the LifeBand has taken up any slack in the bands (see Figure 3-10).

If the patient is not properly aligned, press the Stop/Cancel button, realign the patient, and begin
compressions again with step 1 on page 3-8.

Warning: Do not touch the patient or the LifeBand while the AutoPulse Platform is analyzing the

patient’s size.

Warning: If the Stop/Cancel button is not pressed within 3 seconds, compressions will automatically

begin. Press the Stop/Cancel button to immediately stop the compressions.

3 0 : 2

STOP

A n a l y z i n g P a t i e n t s i z e . . .

D O N O T T O U C H P A T I E N T .

Advertising