Appendix b technical specifications, B.1 patient parameters, B.2 lifeband – ZOLL AutoPulse 100 User Manual

Page 77: B.3 operating parameters b.4 platform physical

Advertising
background image

P/N 12555-001 Rev 3

Page B-1

User Guide

Appendix B Technical Specifications

The specifications provided in this chapter apply to the AutoPulse Resuscitation System Model 100.

B.1

Patient Parameters

The AutoPulse is designed for adults with weight of no more than 300 lbs. (136 kg) with chest
circumference of 29.9 to 51.2 in. (76 to 130 cm) and chest width of 9.8 to 15 in. (25 to 38 cm).

B.2

LifeBand

The latex-free LifeBand is for single-patient use only. The LifeBand consists of a cover plate and two
bands integrated with a patient liner and compression pads with a Velcro

®

fastener.

B.3

Operating Parameters

B.4

Platform Physical

Warning: The AutoPulse System is intended for use on adults, 18 years of age or older.

Table B-1

Operating Parameters

Category

Specifications

Chest displacement

Equal to 20% reduction in anterior-posterior chest depth.

Physiological duty cycle

50 ± 5%.

Compression rate

80

±

5 compressions per minute.

Compression modes
(user selectable)

30:2 (30 compressions with two 1.5 second ventilation pauses)

15:2 (15 compressions with two 1.5 second ventilation pauses)

Continuous compressions

Ventilation pause
(30:2 and 15:2 mode)

Two pauses of 1.5 seconds.

Table B-2

Physical Specifications

Category

Specifications

Manufacturer

ZOLL Circulation, Inc.

Size (L

×

W

×

H)

32.5 in. by 17.6 in. by 3.0 in. (82.6 cm by 44.7 cm by 7.6 cm).

Weight (excluding AutoPulse
Battery)

20.5 lbs. (9.3 kg).

Display

Dot matrix liquid crystal display (LCD), actively backlit, adjustable contrast.