Aligning the LifeBand

8.

Close the LifeBand around the patient's chest.

Note:

AED or defibrillation pads are not components of the LifeBand.

To properly align the two sides of the LifeBand:

a)

Place band

on top of patient’s chest (see Figure 3-7).

b)

Locate mating slot of band

over the alignment tab

(see Figure 3-7).

c)

Press the bands together to engage and secure the Velcro

®

fastener (see Figure 3-8).

d)

Lift up the LifeBand to its fullest extension, ensuring that the side bands are at a 90 degree
angle to the platform, that they are not twisted and that there are no obstructions.

e)

Center the LifeBand on the patient's chest, placing it such that its center is over the area
upon which manual compressions are conducted.

Note: If the bands cannot be closed, use manual CPR.