ZOLL AutoPulse 100 User Manual

Page 68

Advertising
background image

Page 4-4

P/N 12555-001 Rev 3

User Guide

3.

Inspect the battery compartment for foreign substances, be sure it is free and clear for battery
insertion. Remove any foreign matter or debris.

4.

Check the vent openings on the bottom-side of the platform, ensure they are free and clear to allow
air flow. Remove any foreign matter or debris that may block the flow of air.

5.

Check the shaft opening where the belt clip is inserted for foreign substances. Blow out the opening
using compressed air or micro-duster. Do not insert tools or other objects into the opening.

6.

Remove the Battery from the AutoPulse, and place it in the AutoPulse Multi-Chemistry Battery
Charger for charging. Place the spare Battery into the AutoPulse. Take a fully charged Battery from
the Charger, and use as a spare Battery.

7.

Perform AutoPulse Self Test - Install a LifeBand and power on the AutoPulse. On Power up, all
LEDs on the display will momentarily light up, then only the Green POWER LED will remain lit.
The Battery Charge Status Icon on the AutoPulse User Control Panel should also be visible with 4
bars, if not replace Battery with a fully charged Battery from the Multi-Chemistry Battery Charger.
If the AutoPulse Platform red ALERT LED remains on, refer to the Chapter 5, “Troubleshooting
Procedures” in this guide. If the RED LED can not be resolved, contact ZOLL.

Inspect the Battery in accordance with the inspection checklist provided below. Follow the
recommended actions as appropriate. The Battery is automatically checked each time the device is
powered on.

1.

Inspect the Battery, including the connector for physical damage that would preclude its insertion
in either the platform or the charger. Contact ZOLL as needed.

2.

Inspect the Battery for cracks in the Battery case exposing internal components. If the Battery is
damaged, do not attempt to place the Battery into the AutoPulse- this can cause damage to the
internal connector of the AutoPulse. If damaged, do not use.

Inspect the Multi-Chemistry Battery Charger in accordance with the inspection checklist provided
below. Follow the recommended actions as appropriate.

1.

Inspect the Multi-Chemistry Battery Charger for physical damage. Contact ZOLL as needed.

2.

Inspect the battery compartments for foreign substances, be sure they are free and clear for battery
insertion. Remove any foreign matter or debris.

If the product malfunctions or fails any of these inspections, contact ZOLL at +1.800.348.9011 or
+1.978.421.9655.