ZOLL AutoPulse 100 User Manual

Page 22

Advertising
background image

Page 1-10

P/N 12555-001 Rev 3

User Guide

To exchange Batteries:

1.

Press the Stop/Cancel button.

2.

Press the ON/OFF button.

3.

Remove the Battery (refer to Section 2.2, “Battery Installation and Removal” for more information).

4.

Install the fully-charged Battery (refer to Section 2.2, “Battery Installation and Removal” for more
information).

5.

Resume chest compressions (refer to Section 3.2, “Starting Chest Compressions” for more
information).

Table 1-2

Battery Charge Status Indicator Specifics

Battery Charge Icon Bars Showing

Charge Level

No bars showing.

The Battery has been depleted. Replace the Battery
immediately.

One bar showing.

The Battery’s capacity is less than one-third of a full charge.
Be prepared to exchange this Battery with a fully-charged
Battery.

Two bars showing.

The Battery’s capacity is between 33% and 66% of a full
charge.

Three bars showing.

The Battery’s capacity is between 66% and 100% of a full
charge.

All bars showing.

The Battery is fully charged.

Advertising