2 cleaning the autopulse platform, 3 storing the autopulse platform, 4 maintenance – ZOLL AutoPulse 100 User Manual

Page 67

Advertising
background image

P/N 12555-001 Rev 3

Page 4-3

User Guide

5.

To remove a Battery from the Multi-Chemistry Battery Charger, grip the Battery while pulling the
finger latch out (Li-Ion Battery) or pressing up on the finger latch (NiMH Battery) to disengage the
locking bar. Pull the Battery straight out until it fully clears the charging bay.

Note:

Newly-charged Batteries can be warm to the touch. This is a consequence of normal operation.

4.2

Cleaning the AutoPulse Platform

1.

Remove and dispose of the LifeBand.

2.

Wipe all the surfaces of the AutoPulse Platform free of foreign matter and spills with a disinfectant
or bactericidal wipe. Check the vents to ensure that they are free and clear of any obstructive matter.

Caution: Do not submerge the AutoPulse in liquid.

Caution: Do not autoclave the AutoPulse Platform, the LifeBand, or the AutoPulse Power System.

3.

Ensure that the AutoPulse is dry before storing.

4.3

Storing the AutoPulse Platform

1.

The AutoPulse should be powered down before it is stored.

Note:

The AutoPulse will automatically power down after 10 minutes of inactivity.

2.

The AutoPulse should be returned to a Carry Case, Hygiene Barrier or Transporter before being
stored. To purchase a Carry Case, Hygiene Barrier or Transporter, please contact your ZOLL
representative by calling +1.800.348.9011 or +1.978.421.9655.

3.

The AutoPulse should be stored in a cool, dry place.

4.

Both the Li-Ion and NiMH Batteries should be stored in the Multi-Chemistry Battery Charger unless
the Battery is deployed in an AutoPulse.

4.4

Maintenance

The AutoPulse System has no user-serviceable parts, nor does the device require regular maintenance.
There are no components that require calibration. Users should periodically inspect the AutoPulse
system to ensure the device’s functionality. In case of repair or service, contact ZOLL at
+1.800.348.9011 or +1.978.421.9655.

Inspect the AutoPulse as described below and follow the recommended actions as appropriate. The
AutoPulse conducts a self-test each time the device is powered on. The self-test is designed to perform
a variety of internal system self-checks for the platform components, and is fully automated. A
recommended AutoPulse Daily Checklist is located in Appendix A.

1.

Inspect the platform for physical damage, including cracks, tears and missing or broken pieces.
Contact ZOLL as needed.

2.

Remove the Battery from the AutoPulse.

Advertising