3 battery charge status – ZOLL AutoPulse 100 User Manual

Page 21

Advertising
background image

P/N 12555-001 Rev 3

Page 1-9

User Guide

1.4.2.8 Power (Green LED)

The green Power LED lights whenever the AutoPulse is powered on and able to respond to user input.

1.4.2.9 Alert (Red LED)

The red Alert LED lights whenever a user advisory, fault or system error condition exists for the
AutoPulse. For a list of advisory, fault and error conditions, refer to Chapter 5, “Troubleshooting
Procedures”.

1.4.3

Battery Charge Status

The User Control Panel displays the battery charge status. The Battery charge status icon only appears
when the AutoPulse is powered up.

Figure 1-5

Low Battery Warning

When five minutes of active operation remain on a Battery, the User Control Panel will give a “Low
Battery” indication (see Figure 1-5). The “Low Battery” indication will remain on until the Battery is
replaced or depleted. The Low Battery warning display will be accompanied with an audio warning of
four rapid beeps which will be followed by two beeps every 30 seconds until the battery is replaced or
depleted. It is recommended that, if available, a fully-charged Battery be exchanged for the Battery with
the low charge.

Indicates the level of charge of the Battery. A graphic battery icon indicating four
proportional levels of battery charge is displayed.

3 0 : 2

LOW BA TTERY

Advertising