ZOLL AutoPulse 100 User Manual

Page 51

Advertising
background image

P/N 12555-001 Rev 3

Page 3-11

User Guide

Figure 3-11 Chest Compression Display Panel Screen

5.

Depending on the Mode setting in Administrative Menu (refer to section Section 2.3,
“Administrative Menu: User Pre-set Options,” on page 2-11), the AutoPulse will perform 30:2, 15:2
or Continuous compressions. In 30:2 mode it performs 30 compressions and then pauses for three
seconds to permit the user to ventilate the patient before automatically resuming compressions (see
Figure 3-12). In 15:2 mode it performs 15 compressions and then pauses for three seconds to permit
the user to ventilate the patient before automatically resuming compressions (see Figure 3-12). In
Continuous mode it performs continuous compressions. If 30:2 on-the-fly mode switching has been
enabled (in the Mode setting within the Administrative Menu) then the AutoPulse will perform in
the mode (either 30:2 or Continuous) that was used last until powered down; on power up 30:2 will
be the initial selection. If 15:2 on-the-fly mode switching has been enabled (in the Mode setting
within the Administrative Menu) then the AutoPulse will perform in the mode (either 15:2 or
Continuous) that was used last until powered down; on power up 15:2 will be the initial selection.

In 30:2 mode three audio cue tones will sound prior to the ventilation pause: one during each of the
28th, 29th, and 30th compressions. In 15:2 mode the three audio cue tones will sound prior to the
ventilation pause during each of the 13th, 14th, and 15th compressions. In Continuous mode, an
audio cue tone for ventilation will sound 8 times per minute. The tones can be temporarily disabled
(and re-enabled) by pressing the Tone Mute Button (refer to Section 1.4.2.6, “Tone Mute Button,”
on page 1-7).

At the initiation of compressions, the counter at the right center of the display panel screen will be
set to 00:00 and will automatically begin recording the elapsed time until the Stop/Cancel button is
pressed. The format of the counter is minutes:seconds. When the Stop/Cancel button is pressed, the
counter will immediately reset to zero and begin recording the elapsed "no-flow" time. The counter
will reset to zero when chest compressions are started again.

3 0 : 2

0 0 : 1 7

STOP

S w i t c h t o
CONTINUOUS

COMPRESSING ...

Advertising