Accuracy (combined with axf remote flowtube), Caution – Yokogawa AXFA14G/C User Manual

Page 105

Advertising
background image

IM 01E20C02-01E

11-3

11. OUTLINE

Accuracy (Combined with AXF Remote Flowtube)

Pulse Output:

PFA/Ceramics Lining:

Size mm

(in.)

V

Ͻ 0.3 (1)

Ϯ1.0 mm/s

Ϯ0.5 mm/s

Ϯ0.5 mm/s

0.3

Ϲ V Ϲ 10

(1) (33)

V Ͻ 0.15 (0.5)

V Ͻ 0.15 (0.5)

0.15

Ϲ V Ϲ 10

(0.5) (33)

0.15

Ϲ V Ϲ 10

(0.5) (33)

V Ͻ0.15 (0.5)

0.15

Ϲ V Ͻ1

(0.5) (3.3)

Ϯ0.35% of
Rate

Ϯ0.5 mm/s

Ϯ0.35% of
Rate

Ϯ0.35% of
Rate

2.5 (0.1)
to
15 (0.5)

250 (10)
to
400 (16)

25 (1.0)
to
200 (8.0)

Flow Velocity

V m/s (ft/s)

Flow Velocity

V m/s (ft/s)

Standard

Accuracy

(Calibration

code B)

High grade

Accuracy

(Calibration

code C)

Ϯ0.18% of Rate

Ϯ0.2mm/s

1

Ϲ V Ϲ10

(3.3) (33)

Ϯ0.2% of Rate

T01.EPS

Enhanced dual frequency excitation (Option code HF2):
Standard accuracy

Ϯ 1 mm/ s

Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)

1.2

1.0

0.8

0.6

0.4

0.35

0.2

0

0

1

2

6

8

10

V[m/s]

High grade Accuracy

High grade Accuracy

Standard Accuracy

F01.EPS

Pulse
Output
Accuracy

% of Rate

Size 25 mm (1.0 in.) to 400 mm (16 in.)

1.2

1.0

0.8

0.6

0.4

0.35

0.2

0

0

1

2

6

8

10

V[m/s]

F02.EPS

Pulse
Output
Accuracy

% of Rate

Standard Accuracy

High grade Accuracy

Polyurethane Rubber/Natural Soft Rubber/EPDM
Rubber Lining

Size mm (in.)

V

Ͻ 0.3 (1.0)

Ϯ1.0 mm/s

0.3

Ϲ V Ϲ 10

(1.0) (33)

Ϯ0.35% of Rate

25 (1.0) to 400 (16)

Flow Velocity

V m/s (ft/s)

Standard Accuracy

(Calibration code B)

T02.EPS

Enhanced dual frequency excitation (Optional code HF2):
Standard accuracy

Ϯ 1 mm/ s

Size 25 mm (1.0 in.) to 400 mm (16 in.)

1.2

1.0

0.8

0.6

0.4

0.35

0.2

0

0

1

2

6

8

10

V[m/s]

High grade Accuracy

High grade Accuracy

Standard Accuracy

F01.EPS

Pulse
Output
Accuracy

% of Rate

Current Output “

”: Pulse output accuracy plus Ϯ0.05% of Span

Repeatability:

Ϯ0.1% of Rate (V м 1 m/s (3.3 ft/s))

Ϯ0.05% of Rate Ϯ0.5 mm/s (V < 1 m/s (3.3 ft/s))

Maximum Power Consumption:

Combined with AXF Remote Flowtube: 12 W

Insulation Resistance(*1) :

Between power supply terminals and ground
terminal: 100M

Ω at 500V DC

Between power supply terminals and input/output/
excitation current terminals : 100M

Ω at 500V DC

Between ground terminal and input/output/excitation
current terminals: 20M

Ω at 100V DC

Between input/output/excitation current terminals
:20M

Ω at 100V DC

Withstand Voltage(*1) :

Between power supply terminals and groundterminal:
1390V AC for 2 seconds
Between power supply terminals and input/output
terminals: 1390V AC for 2 seconds
Between excitation current terminal and ground terminal:
160V AC for 2 seconds
Between excitation current terminal and input /output
terminals: 200V AC for 2 seconds

CAUTION

*1: When performing the Insulation Resistance Test or the

Withstand Voltage Test, please obey the following
caution.

• Following the relevant test, wait for more than 10

seconds after the power supply has been turned off
before removing the cover.

• Remove all wires from terminals before testing.
• When the power terminal has a lighting protector

(optional code A), remove the short bar at the ground
terminal.

• After testing, be sure to discharge by using a resis-

tance and return all wires and the short bar to its
correct position.

• Screws must be tightened to a torque of 1.18 N-m or more.
• After closing the cover, the power supply can be restored.

Advertising
This manual is related to the following products: