4 wiring connections, 1 removing cover, 2 terminal configuration – Yokogawa AXFA14G/C User Manual

Page 16: Wiring connections -4, Removing cover -4, Terminal configuration -4

Advertising
background image

IM 01E20C02-01E

4-4

4. WIRING

(3) Conduit Wiring

When wiring the conduits, pass the conduit through the
wiring connection port, and utilize the waterproof
gland to prevent water from flowing in. Place the
conduit pipe on an angle as shown in Figure 4.3.4.
Install a drain valve at the low end of the vertical pipe,
and open the valve regularly.

Drain valve

F0407.EPS

Figure 4.3.4

Conduit Wiring

4.4 Wiring Connections

4.4.1

Removing Cover

Loosen cover locking screw 2 clockwise using a
hexagonal wrench (nominal size 3 mm) to unlock the
cover. (Upon shipment from the manufacturing plant,
the cover is unlocked.) Hold the flowmeter with your
hand and remove the cover by turning it in the direc-
tion of the arrow as shown below.

F0408.EPS

Cover locking screws

Figure 4.4.1

Removing the Terminal Box Cover

4.4.2

Terminal Configuration

When the cover is removed, the connection terminals
will be visible. The terminal configuration labels are
attached in the locations shown in Figure 4.4.2.

F0409.EPS

Figure 4.4.2

Terminal Layout Labels Position

The description of the terminal symbols is shown in
Table 4.1.1.

For F

OUNDATION

Fieldbus protocol, please refer to IM

01E20F02-01E.

For PROFIBUS PA protocol, please refer to IM
01E20F12-01E.

Table 4.1.1 Terminal Symbols

T0401.EPS

Flow singal
input

Excitation current
output

Power supply

Current output
4 to 20mA DC

Pulse output/
Alarm output/
Status output

Alarm output/
Status output/
Status input

Functional

grounding

Terminal
Symbols

Description

EX1
EX2

C

SA

A
B

SB

Terminal
Symbols

Description

N/–

L/+

I+
I–

DO+
DO–

DIO+
DIO–

Protective grounding
(Outside of the terminal)