2 fm, Fm -2 – Yokogawa AXFA14G/C User Manual

Page 113

Advertising
background image

IM 01E20C02-01E

12-2

12. EXPLOSION PROTECTED TYPE INSTRUMENT

(3) Installation

WARNING

• All wiring shall comply with local installation

requirements and local electrical code.

• In hazadous locations, the cable entry devices

shall be of a certified ATEX flameproof type,
suitable for the conditions of use and correctly
installed.

• Unused apertures shall be closed with suitable

flameproof certified blanking elements. (The
plug attached is flameproof certified.)

(4) Operation

WARNING

• After de-energizing, delay 20 minutes before

opening.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(5) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the certification.

(6) Data Plate

Made in

mA (0-750

)

VDC 0.2A MAX.

˚C

COMB.NO.

VDC 12W

VAC 50/60Hz 30VA 12W

0344

2G
1D

TOKYO 180-8750 JAPAN

WARNING User’s Manual

No.: KEMA 03ATEX2435
EEx d C T6
ENCLOSURE: IP66, IP67
Tamb: –40 TO +60

°C

MAX.SURFACE TEMP.: T75

°C

FOR DUST-PROOF

*2)

*3)

MODEL

SUFFIX

STYLE

SUPPLY

OUTPUT

AMB.TEMP.

TAG NO.

NO.

MODEL: Specified model code
SUFFIX: Suffix codes of the model code
STYLE: Specified style code
SUPPLY: Power supply voltage of apparatus
OUTPUT: Output signal of apparatus
AMB. TEMP., Tamb: Ambient temperature
No.: Manufacturing serial number *1)

CE: CE marking
II 2G: Group II Category 2 Gas atmosphere
II 1D: Group II Category 1 Dust atmosphere
No.: KEMA 03ATEX2435:

EC Type Examination certificate number

EEx d IIC T6: Protection type and temp. class
ENCLOSURE: Enclosure protection code

WARNING

: Warning to apparatus

YOKOGAWA

᭜ TOKYO 180-8750 JAPAN :

Name and address of manufacturer

*1) The third figure from the last shows the last one

figure of the year of production. For example, the
year of production of the product engraved as
follows is year 2005.

No. S5EA05158 545

Produced in 2005

*2) The identification number of the notified body:

0344 KEMA Netherland

*3) The product-producing country

12.2 FM

(1) Technical Data

Applicable Standard:

FM3600, FM3615, FM3810, ANSI/NEMA 250

Explosion proof for Class I, Division 1, Groups A,

B, C & D.

Dust-ignition proof for Class II/III, Division 1,

Groups E, F & G.

“SEAL ALL CONDUITS WITHIN 18 INCHES”
“WHEN INSTALLED IN DIV. 2, SEALS NOT
REQUIRED”

Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: NEMA 4X
Temperature Code: T6
Ambient Temp.: –40

°C to +60°C (–40°F to +140°F)

(2) Installation

WARNING

• All wiring shall comply with National Electrical

Code ANSI/NFPA 70 and Local Electrical
Code.

• In hazardous locations, wiring to be in conduit

as shown in Figure 12.2.1.

• When installed in Division 2, “SEALS NOT

REQUIRED”

Advertising
This manual is related to the following products: