Yokogawa AXFA14G/C User Manual

Page 39

Advertising
background image

IM 01E20C02-01E

6-10

6. PARAMETER DESCRIPTION

T0610-2.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

Power Synch
(POWER SYNCH)

No
Yes

W

2

Memo 3
(MEMO 3)

R

Memo 1
(MEMO 1)

Power Frequency
(POWER FREQ)

Memo 2
(MEMO 2)

W

47.00 to 63.00

Hz

ASCII 16 characters

Yes

ASCII 16 characters

W

ASCII 16 characters

R/W

W

Software Rev No
(SOFTWARE REV)

Good
Error

R

J50

J60 —

(SELF CHECK)

See “6.5 Alarm Functions”.

Memo field

Memo field

Memo field

Software revision number

J42

Selects whether or not the internal
frequency is to be synchronized
with the power supply frequency.

50.00

J41

J30

J31

J40

Displays the power-supply
frequency (for Power Synch =
“Yes”), or sets the power-supply
frequency (for Power Synch=“No”).

(10) Item K (Menu K): Diagnostic Function Setting items

Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.

T0611.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

2

2

2

Adhesion Level3
(ADH LEVEL3)

Adhesion Level2
(ADH LEVEL2)

M ohm

Adhesion Level4
(ADH LEVEL4)

Adhesion Level1
(ADH LEVEL1)

Adh Measure Value
(ADH MEAS VAL)

Adhesion Check
(ADHESION CHK)

Diagnosis
(DIAGNOSIS)

0.00 to 100.00

0.00 to 100.00

M ohm

M ohm

0.00 to 100.00

M ohm

W

W

W

0.00 to 100.00

No
Yes

W

W

2

Good
Error

R


(SELF CHECK)

K15

K60

K14

Sets the resistance value for adhesion
Level 4 to the electorode.

Sets the resistance value for adhesion
Level 2 to the electorode.

Sets the resistance value for adhesion
Level 1 to the electorode.

Sets the resistance value for adhesion
Level 3 to the electorode.

R

K13

Displays the resistance value for
adhesion to the electrode.

See “6.5 Alarm Functions”.

1.00

K12

0.10

0.50

2

M ohm

3.00

K00

K10

K11

Selects whether or not to perform diagnosis
of adhesion to the electrode.

No

(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items

Menu M contains items that are relevant to automatic zero adjustment.

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

T0612.EPS

Auto Zero Exe
(AUTOZERO EXE)

R/W

Magflow Zero
(MAGFLOW

3

-99.999 to 99.999

Adjustment
(ADJUSTMENT)

0.000

No Execution
Execution

W

Good
Error

R


(SELF CHECK)

Displays the result of the automatic
zero adjustment, or sets the zero point.

See “6.5 Alarm Functions”.

No Execution

Selects whether or not automatic
zero adjustment is carried out.
Linked with B50.

M10

M11

M00

M60

Advertising
This manual is related to the following products: