6 precautions for theaxf integral flowmeter, Precautions for the axf integral flowmeter -40, 6 precautions for the axf integral flow- meter – Yokogawa AXFA14G/C User Manual

Page 69

Advertising
background image

IM 01E20C02-01E

6-40

6. PARAMETER DESCRIPTION

6.6 Precautions for the

AXF Integral Flow-
meter

Although the AXF integral flowmeter’s software
configurations are the same as the AXFA14, care must
be taken in setting the following parameters:

[C31: Nominal Size Unit] Setting of the nominal
size units
[C32: Nominal Size] Setting of the nominal size

NOTE

The nominal size unit, nominal size and meter
factor have been properly set upon shipment
from the manufacturing plant. Do not change
these parameters.

[C21: Low MF] Setting of the low-frequency meter
factor
[C22: High MF] Setting of the high-frequency meter
factor
[C23: Low MF (EDF)] Setting of the low-frequency
meter factor for enhanced DF (i.e., enhanced dual
frequency excitation)
[C24: High MF (EDF)] Setting of the high-frequency
meter factor for enhanced DF (i.e., enhanced dual
frequency excitation)

NOTE

Setting the meter factor

F0623.EPS

Meter factors for the AXF integral flowmeter

0038

(1) Confirm that the meter factors indicated on

your AXF integral flowmeter’s data plate has
been set to the parameters.

(2) The meter factors are crucial in ensuring that

the electromotive force is correctly in propor-
tion to the flow velocity and are determined
at the manufacturing plant by actual-flow
calibration.

NOTE

For the hardware configuration of the AXF
integral flowmeter, see the user’s manual of the
AXF Integral Flowmeter [Hardware Edition] (IM
01E20D01-01E).