6. PARAMETER DESCRIPTION

Parameter setting sequence

(for automatic multiple ranges switching)

F0605.EPS

F20: DO Function
Select a function.

F21: DIO Function
Select a function.

F23: DIO Active Mode

Select whether DIO output
is to be “Closed/Short Act”
or “Open/Open Act”.

Select a function (Note 1)
No Function: Output is stopped (other functions

may be selected)

Fwd/Rev Rngs(O): Forward/reverse flow rate

measurement

Auto 2 Rngs(O): Automatic 2-ranges switching
Auto 3 Rngs(O): Automatic 3-ranges switching
Auto 4 Rngs(O): Automatic 4-ranges switching

Refer to Table 6.4.1 and set
F20 or F21.
Note 1: "DIO (O)" indicates

DIO function is used for output.

B23: Flow

Span

Forward No.1 range

F30: Forward

Span 2

Forward No.2 range

F31:Forward

Span 3

Forward No.3 range

F32:Forward

Span 4

Forward No.4 range

F33:Reverse

Span 1

Reverse No.1 range

F34:Reverse

Span 2

Reverse No.2 range

F22: DO Active Mode

Select whether DO output
is to be “Closed (On) Act”
or “Open (Off) Act”.

If “instantaneous flow rate % (FR)” is selected for
Display Mode and A10: FLOW RATE (%) is selected
for BRAIN communication, the instantaneous flow rate
% and following symbols will be displayed only for
multiple ranges and forward/reverse flow measurement.
Forward No.1 range: [F1]

Reverse No.1 range: [R1]

Forward No.2 range: [F2]

Reverse No.2 range: [R2]

Forward No.3 range: [F3]
Forward No.4 range: [F4]

Set the spans for the ranges to be used.
No.1 range

ϹNo.2 rangeϹNo.3 rangeϹ

No.4 range

10% of No. 1 range
(set using F40)

10% of No. 2 range
(set using F40)

10% of No. 3 range
(set using F40)

Input

100%

No. 1 range

No. 2 range

No. 3 range

No. 4 range

Output

Reverse

Forward

Minimum span % within multiple and
forward/reverse (set using F41)

F0606.EPS

Figure 6.4.1 Multiple Ranges and Hysteresis Widths

NOTE

For more details regarding the setting of hyster-
esis width, refer to the description of setting
parameter for F40: Auto Range Hys and F41:
Bi Direction Hys
.