3 self-diagnostic, 4 hart specific functions, Self-diagnostic -4 – Yokogawa AXFA14G/C User Manual

Page 80: Hart specific functions -4

Advertising
background image

IM 01E20C02-01E

8-4

8. OPERATION VIA HART COMMUNICATOR

8.4.3

Self-diagnostic

The self-diagnostic function of the AXFA14 is ex-
plained in Section 6.5 “Alarm Functions.”
The HART configuration tool can be used to run
self-diagnostics on a device and check for incorrect
data settings.

(1) Using DD

The Self test and Status commands are available for
self-diagnostics. When Self test is run, the display unit
shows an error code and alarm message if the device
detects any illegal parameter settings or functional
faults.

• Procedure to call up the Self test display

Device Setup

→ Diag/Service → Test → Self test

If no error is detected, “Self test OK” is displayed on
the configuration tool.
If the specific diagnostic item is known for the check,
you can directly call up the item by using the Status
command.
The status is categorized from 1 to 7.
Show an example below to confirm the status of Status
group 1.

• Procedure to call up the Status display

Device Setup

→ Diag/Service → Status → Status group 1

If no error is detected, “Off” is displayed on the
configuration tool.
If there is an error, “On” is displayed on the configura-
tion tool, and a countermeasure for that error is
necessary.

Example of display:

Span > 10 m/s

On

Span < 0.3 m/s

Off

TTL > 10000 p/s

Off

TTL < 0.0001 p/s

Off

4-20 Lmt Err

Off

The HART configuration tool diagnoses at each
communication.
When an improper operation is performed, the error
message is displayed.

(2) Using DTM

The Device Status commands are used for
self-diagnostics. When Device Status is run, the
display unit shows an error code and alarm message if
the AXFA14 detects any illegal parameter settings or
functional faults.

• Procedure to call up the Device Status display

Device Status

If no error is detected, “Status: Normal” is displayed
on the configuration tool.
If the specific diagnostic item is known for the check,
you can directly call up the item by using the Diagnos-
tic List in the Device Status display.
The Diagnostic List is categorized to Device Status,
Hardware Failure, Transducer Status, Diag Status, and
Configuration.
If no error is detected, color symbol which shows
Normal State is displayed on top of the error message.
If color symbol which shows Error State is displayed,
there is an error and a countermeasure for that error is
necessary.

The HART configuration tool diagnoses at each
communication.
When an improper operation is performed, the error
message is displayed.

8.4.4

HART Specific Functions

᭿ Trim Analog Output
This function is used to adjust the analog output at
4 mA and 20 mA with the D/A trim or Scaled D/A trim.

(1) D/A trim

Connect a calibration digital ammeter, and then
enter the read value of the ammeter for each output
of AXFA14.

• Procedure to call up the D/A trim display

DD

Device Setup

→ Diagnosis/Service →

Adjustment

→ D/A trim

DTM

Calibration

→ D/A trim

T0808.EPS

(2) Scaled D/A trim

The output is adjusted by using a voltmeter or
other types of meters with 0 to 100% scale.
Example using a voltmeter:

4 mA DC

→ 1 V

20 mA DC

→ 5 V

Connect this meter, and enter the read value of the
meter for each output of AXFA14.

• Procedure to call up the Scaled D/A trim

display

DD

Device Setup

→ Diagnosis/Service

→ Calibration

→ Analog output trim →Scaled D/A trim

DTM

Calibration

→ Scaled D/A trim

T0809.EPS