Yokogawa AXFA14G/C User Manual

Page 54

Advertising
background image

IM 01E20C02-01E

6-25

6. PARAMETER DESCRIPTION

Multiple Ranges Setting 2:

Multiple ranges switching via external
status input

᭿ For both the forward and reverse directions, switch-

ing of up to two ranges can be carried out based on
a status input; however, switching between direc-
tions is not possible. Switching between forward
and reverse ranges is carried out automatically only
when the flow direction reverses.

᭿ DIO terminal is used for multiple ranges switching.

For more details, refer to Table 6.4.2: Multiple
Ranges Switching via External Status Input.

Multiple Ranges Switching via External Status
Input

Operations are performed in accordance with the
following table when the active mode has been set to
“Closed (On) Act” using F22: DO Active Mode or
“Closed/Short Act” using F23: DIO Active Mode.
Operating patterns are reversed when the active mode
has been set to “Open (Off) Act” using F22: DO
Active Mode
or “Open/Open Act” using F23: DIO
Active Mode
.

Table 6.4.2 Multiple Ranges Switching via External Status

Input

T0628.EPS

*: “No Function” is the default value. Only DIO terminal is used for

dual ranges; accordingly, the unused DO terminal remains at the
default value (i.e., No Function) and can therefore be used for
other functions.

*: “DIO (I)” indicates DIO function is used for input.

Parameter setting

Function

F20

Status output

DIO (O) DIO (I)

DIO

Open
Short

Open
Short
Open
Short

Open
Short






DO


Open
Open

Closed
Closed

Open

Closed

No
Function

Fwd/Rev
Rngs(O)

Ext 2
Answer(O)

F21

Ext 2
Ranges(I)

Ext 2
Ranges(I)

Ext 2
Ranges(I)

External forward
2 ranges

External forward/
reverse 2 ranges

External forward
2 ranges, with
answerback

Forward 1 range
Forward 2 range

Forward 1 range
Forward 2 range

Reverse 1 range
Reverse 2 range

Forward 1 range
Forward 2 range

Parameter setting sequence

(for multiple ranges switching via external
status input)

F20: DO Function
Select a function.

F21: DIO Function
Select a function.

F23: DIO Active Mode
Select whether DIO input is
to be “Closed/Short Act”
or “Open/Open Act”.

Select a function (Note 1)
No Function: Function is stopped (other functions may be selected)
Fwd/Rev Rngs(O): Forward/reverse flow rate measurement
Ext 2 Answer(O): Answerback; 2-ranges switching via external

status input

Ext 2 Ranges(I): 2-ranges switching via external status input

Refer to Table 6.4.2 and set F20 or F21.
Note 1: "DIO (O)" indicates DIO function

is used for output and "DIO (I)"
for input.

F0607.EPS

F22: DO Active Mode
Select whether DO output is
to be “Closed (On) Act” or
“Open (Off) Act”.

F34: Reverse

Span 2

Reverse No.2 range

F33: Reverse

Span 1

Reverse No.1 range

F30: Forward

Span 2

Forward No.2 range

B23: Flow

Span

Forward No.1 range

If “instantaneous flow rate % (FR)” is selected for Display
Mode and A10: FLOW RATE (%) is selected for BRAIN
communication, the instantaneous flow rate % and following
symbols will be displayed only for multiple ranges and
forward/reverse flow measurement.
Forward No.1 range: [F1]

Reverse No.1 range: [R1]

Forward No.2 range: [F2]

Reverse No.2 range: [R2]

Set the spans for the ranges to be used.
No.1 range

ϹNo.2 range

[F40: Auto Range Hys] Setting of automatic range-
switching hysteresis width
Automatic switching takes place for multiple range
switching when 100% of the range is exceeded, and
this parameter allows a hysteresis width to be set for
this switching.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.

[F41: Bi Direction Hys] Setting of forward/reverse
flow measurement hysteresis width
This parameter sets the hysteresis for forward/reverse
flow rate measurement as a % value of the minimum
flow span.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.