4. WIRING

(2) Connection with the Remote Flowtube

(Explosion proof Type, Size 2.5 to 400 mm
(0.1 to 16 in.))

In case of explosion proof type for ATEX, FM, CSA,
IECEx and TIIS certification, connect wiring as shown
in the figure below.

In case of the explosion proof type, the protective
grounding

of remote flowtube must be connected to

a suitable IS grounding system. In that case,
(functional grounding terminal) need not be connected.

AXFC dedicated
signal cable

EX2

EX1

A

B

C

Excitation cable

EX1

EX2

C

SA

A

B

SB

F0414.EPS

Converter

Remote

flowtube

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

AXFA14 Converter

Remote flowtube

Figure 4.4.6

Wiring Diagram

4.4.7

Connecting to External Instru-
ments

WARNING

Before wiring with external instruments, be sure
to turn off the power supply for AXFA14
converter and any external instruments.

Connect the AXFA14 terminal to external instruments,
giving attention to the following points.

For F

OUNDATION

Fieldbus protocol, please refer to IM

01E20F02-01E.

For PROFIBUS PA protocol, please refer to IM
01E20F12-01E.

4 to 20 mA DC Current Output

Resistive load max. 750

Ω.

(When using BRAIN/ HART communication,
more than 250

Ω, less than 600 Ω).

Receiver

Instrument

AXFA14

l+

l-

F0415.EPS

Figure 4.4.7

4 to 20 mA DC Output Connection

Pulse Output

IMPORTANT

• As this is a transistor contact (insulated type),

give attention to proper voltage and polarity
when wiring.

• Do not apply a voltage larger than 30V DC or a

current larger than 0.2A in order to prevent
damage to the instrument.

• When input filter constant of the electronic

counter is large in relation to the pulse width,
the signal will decrease and the count will not
be accurate.

• If the input impedance of the electronic counter

is large, an induction noise from the power
supply may result in inaccurate counts. Use a
shield cable or sufficiently reduce the input
impedance of the electronic counter within the
electromagnetic flowmeter pulse output specifi-
cation range.

• The active pulse output (Optional code EM)

cannot be used in conjunction with the standard
pulse output.

• When the active pulse output (Optional code

EM) is selected, do not be short-circuit between
the DO+ and DO– terminals to avoid damaging
the instrument.

• When the active pulse output (Optional code

EM) is selected, the range of pulse rate must
be set to 2 pps maximum.

• To avoid communication (BRAIN/ HART)

failure, it is recommended to use the shield
cable.

NOTE

For pulse output from the DO terminals, param-
eters must be set. Refer to “Chapter 6: Param-
eter Description” in this manual.

F0416.EPS

Mechanical Counter

Electronic Counter

Load

Protective diode

30V DC, 0.2A. max

PULSE OUT

PULSE OUT

AXFA14

AXFA14

DO+

DO-

DO+

DO-

Figure 4.4.8

Pulse Output Connection