IM 01E20C02-01E

6-27

6. PARAMETER DESCRIPTION

Output Example 2

The high alarm (H) is set to 80% or more of the flow
rate span; the low-low alarm (LL), to 20% or less.

Settings are:

G10: Low Alarm = -110%

G11: High Alarm = 80%

G12: Low Low Alarm = 20%

G13: High High Alarm = 110%

0

Time (t)

H

DO

DIO

LL

Instantaneous flow rate

20

50

80

100

[%]

Select “H/L Alarm(O)” for F20: DO Function
Select “HH/LL Alarm(O)” for F21: DIO Function
Select “Closed (On) Act” for F22: DO Active Mode
Select “Closed/Short Act” for F23: DIO Active Mode

F0609.EPS

L Alarm set to
-110% indicates
that the alarm is
disabled.

HH Alarm set to 110% indicates
that the alarm is disabled.

High Alarm

Low Low Alarm

Output Example 3

The high alarm (H) is set to 80% or more of the flow
rate span; the high-high alarm (HH), to 90% or more.

Settings are:

G10: Low Alarm = -110%

G11: High Alarm = 80%

G12: Low Low Alarm = -110%

G13: High High Alarm = 90%

0

Time (t)

H

HH

DO

DIO

Instantaneous flow rate

50

80

90

100

[%]

F0610.EPS

High High Alarm

High Alarm

Select “H/L Alarm(O)” for F20: DO Function
Select “HH/LL Alarm(O)” for F21: DIO Function
Select “Closed (On) Act” for F22: DO Active Mode
Select “Closed/Short Act” for F23: DIO Active Mode

L Alarm set to
-110% indicates
that the alarm is
disabled.

LL Alarm set to -110% indicates
that the alarm is disabled.

NOTE

• Although the same items can be selected using

the DO terminal (selected for F20) and the DIO
terminal (selected for F21), output is identical
for both.

• Setting values of -110% and 110% are used to

disable corresponding functions; and accord-
ingly, status output can be customized for
specific purposes.