Query

SScaleMath[,p1]?

Example Display the math channel 015 scale at display

position 1. Display four equally spaced main

scale marks.

SScaleMath,015,1,4

SBarMath

Bar Graph Display (/MT)

Sets the bar graph display of a math channel.

Syntax SBarMath,p1,p2,p3

p1 Channel number

p2 Bar display base position

Lower Lower

Center Center

Upper Upper

p3 Number of scale divisions (4 to 12)

Query

SBarMath[,p1]?

Example Display the computed values of math channel

015 on a bar graph with the center set as the

base position (Center). Display four equally

spaced main scale marks.

SBarMath,015,Center,4

SPartialMath

Partial Expanded Display (/MT)

Sets the partial expanded display of a math channel

waveform.

Syntax SPartialMath,p1,p2,p3,p4

p1 Channel number

p2 Partial expanded On/Off (On, Off)

p3 Partial expanded boundary position [%] (1

to 99)

P4 Partial expanded boundary value

Query

SPartialMath[,p1]?

Example For channel 015 whose measurement range is

0 to 1.0000 V, display the measured value of

0.7500 V at the 50% position.

SPartialMath,015,On,50,7500

Description

You cannot set this on a “Off” channel. p2 is fixed to

Off.

If p2=Off, you cannot set p3 or subsequent

parameters.

P2=On can be specified when the difference between

the span upper and lower limits is 2 digits or greater.

SBandMath

Color Scale Band (/MT)

Sets the color scale band of a math channel.

Syntax SBandMath,p1,p2,p3,p4,p5,p6,p7

p1 Channel number

p2 Color scale band (Off, In, Out)

p3 R value of the color scale band RGB

colors (0 to 255)

P4 G value of the color scale band RGB

colors (0 to 255)

p5 B value of the color scale band RGB

colors (0 to 255)

p6 Upper limit of the color scale band display

(span lower limit to span upper limit)

p7 Lower limit of the color scale band display

(span lower limit to span upper limit)

Query

SBandMath[,p1]?

Example For math channel 015, set a blue band in the

range of -0.5000 to 1.0000.

SBandMath,015,In,0,0,255,5000,

10000

Description

You cannot set this on a “Off” channel. p2 is fixed to

Off.

If p2=Off, you cannot set p3 or subsequent

parameters.

For details on RGB values, see “Description” of the

SColorIO

command.

SAlmMarkMath

Alarm Mark (/MT)

Sets the display of the marker that indicates the specified

alarm position of a math channel.

Syntax SAlmMarkMath,p1,p2,p3,p4,p5,p6,p7,

p8,p9,p10,p11,p12,p13,p14,p15

p1 Channel number

p2 Whether to display the alarm mark on the

scale (Off, On)

p3 Alarm mark type

Alarm Display the default alarm mark

Fixed Display the mark with the

specified color

P4 R value of the RGB mark colors for alarm

1 ( 0 to 255)

p5 G value of the RGB mark colors for alarm

1 ( 0 to 255)

p6 B value of the RGB mark colors for alarm

1 ( 0 to 255)

p7 R value of the RGB mark colors for alarm

2 ( 0 to 255)

p8 G value of the RGB mark colors for alarm

2 ( 0 to 255)

p9 B value of the RGB mark colors for alarm

2 ( 0 to 255)

p10 R value of the RGB mark colors for alarm

3 ( 0 to 255)

p11 G value of the RGB mark colors for alarm

3 ( 0 to 255)

p12 B value of the RGB mark colors for alarm

3 ( 0 to 255)

p13 R value of the RGB mark colors for alarm

4 ( 0 to 255)

p14 G value of the RGB mark colors for alarm

4 ( 0 to 255)

p15 B value of the RGB mark colors for alarm

4 ( 0 to 255)

Query

SAlmMarkMath[,p1]?

Example Display the alarm marks for alarms 1 to 4 of

math channel 015 in fixed colors red, brown,

orange, and yellow, respectively.