SECTION 11: FLOW-SENSE WIZARD

rev. 003 11/04

Measurement interval—Auto Calculated Value: This is the software-
calculated time (seconds) for one complete exchange of water in the
calculated volume at the final pumping rate. In other words, this is the
time required to completely evacuate one volume of the flow cell and
tubing and draw in a new volume. Readings will be taken at this
interval unless you specify a different interval.

To specify a different interval, clear the “Auto” check box as shown
above, and enter the desired measurement interval (seconds).

Stabilized drawdown: Final drawdown from the initial water level
entered previously.

Click or tap Next to continue.

Pumping
Information:
Measurement
Interval

TIP: You may
override the
calculated

measurement interval if you
wish to read parameters at
a different interval than
once per volume exchange.