SECTION 11: FLOW-SENSE WIZARD
rev. 003 11/04
Select Channels &
Stabilization
5. The final screen displays the Low Flow parameters. Choose the
parameters to monitor and the stability range for each parameter.
The percentage range is calculated by the formula:
If no percentage ranges are entered, the software will monitor absolute
values values to determine stabilization. If any percentage range is
entered, the software will monitor percentage change.
Click the Notes button to add comments, if you want. They will be
included in the output report.
TIP: If a
parameter you
wish to monitor is
greyed out, be sure the
sensor is properly installed.
Refer to the Multi-
Parameter TROLL 9000
Operator’s Manual on the
Win-Situ 4 installation CD,
or to installation instructions
shipped with the water
quality sensors.
max - min of last 3 readings
last reading
Ч
100%