KYORITSU 6300 Instruction Manual User Manual

Page 149

Advertising
background image

MODEL6300

13. 10

MODEL6300

Section 13 Specification

Section

13

< MODEL8126 >

Ɣ Rated current

AC 200A rms (AC 283A peak)

Ɣ Output voltage

AC 0 ~ 500mV (AC 500mV/ 200A)

Ɣ Measuring range and accuracy (sine wave input)

Measuring range

Frequency range

Accuracy

50/60Hz ±0.5

㧑rdg±0.1mV

0 ~ 200

40Hz ~ 1kHz

±1.0

㧑rdg±0.2mV

Ɣ Phase characteristic

within ±1deg (2 ~ 200A/ 45~65Hz)

Ɣ General specification

Location for use

In door use, Altitude up to 2000m

Temperature & humidity range
(guaranteed accuracy)

23ºC±5ºC/Relative humidity 85% or less
(no condensation)

Operating temperature &
humidity range

0ºC ~ 50ºC/Relative humidity 85% or less
(no condensation)

Storage temperature &
humidity range:

-20ºC ~ 60ºC/Relative humidity 85% or less
(no condensation)

Max allowable input

AC 200A rms continuously (50/60Hz)

Output impedance

approx. 5

ȍ

Withstand voltage

AC5350V rms (50/60Hz)/ for 5sec.

between jaw matched part and enclosures

between enclosures and output terminal
between jaw matched part and output terminal

Insulation resistance

50M

ȍ or more/ 1000V

between jaw matched part and enclosures
between enclosures and output terminal
between jaw matched part and output terminal

Applicable standard

IEC 61010-2-032

Measurement CATIII. 600V, Pollution degree 2

IEC 61326

Diameter of measured conductor

Max.

ĭ40mm

Cable length

About 3m

Output terminal

MINI DIN 6PIN

Outer diameter

128(L) x 81(

㨃) x 36(D) mm

Weight approx.

260g

Accessories

Instruction manual, Cable maker

Advertising