Preparation for measurement, 1 power cord connection – KYORITSU 6300 Instruction Manual User Manual

Page 21

Advertising
background image

3. 1

MODEL6300

Section 3 Preparation for measurement

MODEL6300


3. Preparation for measurement

This section explains the setting up of the instrument.


3-1 Power cord connection

Ɣ Check the following before connecting the Power cord.

DANGER

Ɣ Use only the Power cord supplied with this instrument.
Ɣ Connect the Power cord mains plug to a mains socket outlet. The mains supply
voltage must not exceed 240V.

WARNING

Ɣ Confirm that the Function switch on the instrument is in the OFF position, and then

connect the Power cord.

Ɣ Connect the Power cord to the instrument first. Cord to be firmly inserted.
Ɣ Never attempt to make measurement if any abnormal conditions are noted,

such as a broken cover and exposed metal parts.

Ɣ When the instrument is not in use, disconnect the Power cord from the outlet.
Ɣ When unplugging the cord from the mains socket outlet, do so by removing the
plug first and not by pulling the power cord.

Ɣ Power cord connection

Follow the procedure below, and connect the Power cord.
(1) Confirm that the Function switch on the instrument is in the OFF

position.

(2) Connect the Power cord to the Power connector on the instrument.
(3) Connect the Power cord plug to a mains socket outlet.

Ɣ Power supply rating

Rated supply voltage: 100 ~ 240VAC

(±10%)

Rated power supply frequency: 45 ~ 65Hz
Max power consumption: 10VA max