3-6 “setting 06” ct ratio – KYORITSU 6300 Instruction Manual User Manual

Page 36

Advertising
background image

MODEL6300

4. 10

MODEL6300

Section 4 Setting

Section

4

‣ ‧•⁈ ‥• •⁈ 
•• ⁈

‣ ⁂․ ⁉ ‣ ⁂‥ ⁉ ‥⁂ ‥⁉ ‥ ⁂…⁉

‧ ″

‣ •″

․ •″

‧• ″

‣•• ″ ․•• ″ ‧• •″ ‣• ••″


4-3-6 “Setting 06” CT ratio


For the detailed information about CT ratio, refer to “5-3 VT/CT” in
this manual.

Setting range

1.00 ~ 10000.0

Default value

1.00

* Using the

Cursor

key whilst in the selection screen, select “Setting 06”.

* Press the

ENTER

key to enter setting change mode.

* The rightmost digit of previous setting (or default value: 1.00) flickers.
* Change the number with the

Cursor

key, and then press the

ENTER

key.











Function of

Cursor

keys

To select the desired digit.

To change the value of selected digit.

When CT ratio is set to other than 1.00, “ ” mark appears on each
measurement screen.

NOTE

* When a value between 0.00 ~ 0.99 is set as a CT ratio, it is automatically
changed to 1.00. In case that a value of 10000 or more is set, it is
automatically changed to 10000.
* When the 3000A clamp sensor is selected at “Setting 04”, and Current range
is set to 3000A at “Setting 03”, the CT ratio should be set to 3.00.
* If the setting of Clamp sensor (“Setting 04”) is changed after setting the CT

ratio, the CT ratio automatically returns to 1.00.

* If a value cannot be set as a CT ratio. A message “ ” is displayed on
the LCD, and the value returns to the previous set value.

(Voltage range “Setting 02”) x 120% x (Current range “Setting 03”) x 120%

x (VT ratio “Setting 05”) x (CT ratio)> 9999G (G=10

9

)

‣ ‧• ⁈ ‥• •⁈ 
 •• ⁈

‣ ⁂ ․⁉ ‣ ⁂‥ ⁉ ‥ ⁂‥ ⁉ ‥⁂ …⁉

‧″

‣ •″

․ •″

‧ •″

‣• •″ ․ •• ″ ‧• •″ ‣ •• •″

Advertising