WIKA Pascal 100 User Manual

Page 35

Advertising
background image

6 Commissioning, operation

WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 35

GB

Selecting RTD the page for the thermoresistance

set up is shown. Fourteen RTD’s types are

available, plus one RTD User (with coefficient defined by the user).

Figure 22

– Rtd type selection


Once the type of thermoresistance connected to the input IN A is identified, the operator must
indicate the measurement mode: 2 wires, 3 wires or 4 wires measurement; the engineering unit
(°C, °F, K) and related number of decimal.


Measurements of RTD - 4 wires
If possible it is recommended to perform RTD measurement with four wires to
eliminate parasite resistances of the connecting cables. All specifications for
thermoresistance aim for four wires measurement.




RTD USER must be selected when the operator needs to measure a non-standard RTD in
agreement with the following formula:
If T≥ 0 °C R(t)= Ro * (1 + A * t + B * T²)
If T< 0 °C R(t)= Ro * (1 + A * t + B * T² + C (t

– 100) * t³)

Figure 23

– RTD parameters setup


Advertising