WIKA Pascal 100 User Manual

Page 8

Advertising
background image

2 Safety

8 WIKA Operating Instruction, Pascal 100 and Pascal 100/IS

GB

2.3

Additional safety instructions for instruments with ATEX approval
only Pascal 100/IS

WARNING!
Non-observance of these instructions and their contents may result in the loss of
explosion protection.

Battery operation:

Use only the rechargeable battery which is supplied by WIKA! Only charge the
battery outside of the hazardous area!

Operation of the serial interface in the hazardous area is prohibited.



Further hazardous area safety instructions!
Observe the operating information and the relevant country-specific regulations
concerning use in hazardous areas (e.g. EN IEC 60079-14).

The Pascal 100/IS intrinsically safe multichannel calibrator has been designed for
use in Ex Hazardous Areas. These are areas where potentially flammable or
explosive vapors may occur. These areas are referred to as hazardous (classified)
locations in the United States, as Hazardous Locations in Canada, as Potentially
Explosive Atmospheres in Europe and as Explosive Gas Atmospheres by most of
the rest of the world. The Pascal 100/IS intrinsically safe multichannel calibrator is
designed as intrinsically safe. This means that connecting the Pascal 100/IS
intrinsically safe multichannel calibrator to equipment that is used within intrinsically
safe circuits will not cause an ignition capable arc as long as the entity parameters
are suitably matched.



Information/Approval for hazardous locations:

WARNING!

Ex Hazardous Areas

An Ex-hazardous area as used in this manual refers to an area made hazardous
by the potential presence of flammable or explosive vapors. These areas are also
referred to as hazardous locations.






WARNING!
Only gauges powered by batteries are approved for use in hazardous locations.
Use only the rechargeable battery which is supplied by WIKA! Only charge the
battery outside of the hazardous area!

II 2G
Ex ib IIC T4 Gb - Tamb: -10°C / 50°C

Advertising