6 Commissioning, operation

WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 59

GB

The instrument shows:


Press “Zero Trim” or “Span Trim”, and the display shows (in this case Span Trim is selected):

The HART instrument is put in fixed mode generation (4 or 20 mA) and the REF channel is
measuring the true analog mA generation. To re-align to the correct 20 mA (or 4 mA) generation
press “Calibrate” button.
If needed follow the same procedure for “Zero Trim”.