6 Commissioning, operation

WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 77

GB

To visualize the report, access to REPORT menu and press VIEW RPT to view the stored reports:

Figure 95

– Display Report list


The above picture shows the list of the existing report: name and status are displayed.
Under the status, the writing AS FOUND is displayed when the calibration data refers to a
calibration process done on an instrument without any adjustment: when a date is available this
means that calibration process is complete (AS FOUND AS LEFT). If the report has never been
run, VOID will be displayed next to the name.
Press the arrows to move the cursor up and down to select the report, press on the right arrow to
visualize the information available in the report.
The menu DELETE allows to cancel the reports: a confirmation is required by the operator before
proceeding

Selecting the name of the REPORT, the SETUP of the report will be displayed:

Figure 96

– Display Report setup


Right arrow

Up & Down
arrows