6 Commissioning, operation
36 WIKA Operating Instruction, Pascal 100 and Pascal 100/IS
GB
Once RTD USER is selected, it is possible to set the parameter of the RTD used.
A
=
X * 10
-3
°C
-1
B
=
X * 10
ˉ7
°C
-2
C
=
X * 10
-12
°C
-3
Ro
=
Resistance @ t = 0,01 °C
X
=
Value
Next step is to define the measurement range. The range selection is necessary during the
calibration procedure to define the relationship between the REF and DUT channels.
Figure 24
– Range setup
Now it is possible to conclude the procedure by pressing the YES button or come back to the
previous step pressing NO.
Figure 25
– Channel setup confirmation
Once the YES key is pressed, the calibrator returns to the initial page where the user can proceed
with the measurement/simulation, or assign another function to a different software channel
following the same procedure.