6 Commissioning, operation

40 WIKA Operating Instruction, Pascal 100 and Pascal 100/IS

GB

The next screen shows the Span parameter (measuring range). For default the sensor has its
maximum range equal to the measuring range (Span). If the Span has to be redefined, or to be
reduced, the same can be changed by using a numeric keyboard. The redefinition of the Span is
used during a calibration for comparison where a relationship between a REF and a DUT channel
has to be done.

Figure 32

– Sensor range setup


Now it is possible to finalize the procedure by pressing the YES button or come back to the
previous step pressing NO.

Figure 33 - Channel setup confirmation


Pressing the key YES, the channel is set up with the selected parameters and the screen with four
channels is displayed again.

Figure 34

– Main display with pressure measurement