background image

6 Commissioning, operation

WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 57

GB

After positioned to the parameter needed to modify press right arrow key, for example on the TAG
parameter:

If desired delete all the characters with backspace and insert the desired. After press ENTER to
store the parameter inside the HART transmitter; the Pascal will return to the previous display.
IMPORTANT NOTE: HART requires that all the characters must be in upper case in order to
be properly stored.
Press NEXT to continue.

The next step shows the span parameter (measuring range). In the default settings, the maximum
range is equal to the measuring range (span). If the span has to be redefined or reduced, it can be
changed by using a numeric keyboard. To access to the keyboard use the arrows displayed on the
right side of the screen. Change of the span is used during a calibration for comparison where a
relationship between a REF and a DUT channel has to be assigned.

Now it is possible to finalize the procedure pressing YES or come back to the previous step
pressing NO.