23. Display ramp values

Ramp values once programmed can be displayed and edited.
Hit Ramp key; Ramp on LED illuminates. Press the Select key (Time) until Set Ramp Value
LED
illuminates.
First programmed time 00:00 and corresponding value are displayed.
Void parameters are marked OFF.
More times and programmed values are displayed with the Set keys.

24. Edit ramp values

To edit a value, select with Select key.
Now edit value with the Set keys.
Accept new value by pressing the Select key once more.
Program more times: refer to para. 1, set ramp values.

25. Start ramp function

Program ramp values as described in 1).
Press Ramp key, Ramp on LED illuminates.
Hit Auto Start Stop key, Auto Start Stop LED illuminates.
The following functions are started:

lift UP

rotation ON

vacuum ON

heater ON
All values programmed for a ramp are controlled automatically.
Display shows time elapsed from START.
Values between times programmed are interpolated.

26. Discontinue ramp function

Ramp function may be discontinued any time.
To discontinue hit the Ramp key, Ramp LED dims.
The very last ramp function values are kept.
A ramp once programmed will be kept in memory even during power-off times and can be
used and edited all the time.

27. Delete ramp

Ramp LED and Set Ramp Value LED must illuminate.
Press both SET keys

at the same time and hold. After 5 seconds, “del” appears for 3 seconds,

a buzzer sounds.
The ramp has been deleted.

With the ramp function activated, it is not possible to also activate the timer function! If the
timer function is activated during the ramp function, the ramp function will be switched off and
the current values of Set Vac, Set Bath, Set Speed, and Set

T auto are accepted as Set-

values. These can then naturally be changed.