DISTILLATION SETTING

1. General

Using rotational evaporators for thermal breakdown of substances needs some basic
considerations about optimizing the distillation process. Parameters stipulated below are to be
considered for making optimal settings.

Rotational speed, evaporator flask

Increasing rotational speed will increase rate of distillation. Shorter distillation time preserves
thermal stability of your substances to be distilled.

Difference between heating bath and vapor temperatures:

This temperature difference should not be less than 20° C to maintain proper rate of distillation.
As long as your substances in distillation allow, you may even chose a higher temperature
difference.

Rule of thumb: doubling actual temperature difference doubles rate of distillation.

Temperature difference between vapor temperature and coolant:
To maintain adequate condensation rates, we recommend a 20° C temperature difference
between the solvent vapor temperature and coolant. A low temperature difference will affect
solvent recovery. For substances with high evaporation heat, you should increase temperature
difference.

2. Select vacuum control mode

2.1.

p

const

vacuum control mode

The p

const

mode includes the following features:

One of the most important modes of control

Requires optional vacuum valve and vacuum pressure sensor

Selective distillation of different substances within the scope of their physical properties

Distillation of solvent mixtures need manual pressure settings

High reproducibility

Requires knowledge of thermodynamic substance data

first choice for all standard processes

For operation & parameter settings refer to page 85

Pressure (set vacuum) and hysteresis

p (set

p) need to be set