5. Install vapor tube and PTFE-seal


Unscrew large-size (1) and small (2) coupling rings. Lock driveshaft by turning drive lock (3) CW to
ease loosening and tightening of large-size ring (1).

Remove PTFE-adapter (4) with built-in PTFE-O-ring (5) and PTFE-seal (6)

Install backing ring (7) and vapor tube (8) in drive head. Put sandwich gasket (9) on vapor tube

coupling flange

Slide small ring (2) over coupling flange of distributor (10). Expand small split ring (11) and

slide over distributor (10) coupling flange, shoulder pointing forward

Secure assembly (10, 2, 11) tight with small coupling ring (2).

Slide PTFE seal (6) on vapor tube (8), sealing lip pointing inward


Never attempt to work with damaged or worn-out PTFE gasket (6). Not reaching vacuum
pressure wanted may damage your rotary evaporator. Inspect PTFE gasket (6) in
regular intervals and replace on condition.

Slide PTFE-adapter (4) with built-in PTFE-O-ring (5) into drive module.

Slide large-size coupling (1) over evaporator flask (12) coupling flange. Expand large

split ring (13) and slide over evaporator flask (12) coupling flange, shoulder pointing
forward. Set flask support tubes to 10 liter or 20 liter mark and lock with knobs (14); (I =
10 l, II = 20 l)