22. Set ramp values

Before programming a new ramp, make sure to have deleted the old one (refer to “delete ramp”,
page 100).

Procedure 1

st

step:

You need to program 00:00 for the first time.
Press the Select key

next to the Display Time

until Set

Ramp Time LED illuminates.
Set Ramp Time LED ON. Display Time only is active,
all other displays are blank.
Display shows time for the present ramp in hh:mm.

The time 00:00 should always be programmed as the
first time (is preset).

Press the Set keys

to alter time setting (forward &

back).
Not pressing a key for 5 seconds, display will return to
Act. Last value entered will be saved.

To program parameter values as a function of times
selected, go to Set Ramp Value LED with the Select key

next to the Display Time

. Set Ramp Value LED

illuminates, Set Ramp Time LED flashes.
Go to parameter desired with Select key

next to the Display Time

, Set LED illuminates:

the following parameters can be programmed:
vacuum (p const mode) Set Vacuum LED illuminates

T auto (T auto mode) Set

T auto LED illuminates

bath temperature

Set Bath LED illuminates

speed

Set Speed LED illuminates

If one or more of the parameters listed above is not to be assigned a ramp function, leave this
parameter as „OFF“.

As shipped, no ramp functions are set. (“OFF“ displayed)
Values are programmed with the Set keys, as described.
Write value in memory by pressing the Select key once more.
Use the same procedure to program the next parameter.

Display

Act Time LED

Set Time LED

Set Ramp Time
LED

Set Ramp Value LED

Select key

Timer LED

Start Stop key

increase

decrase