Carefully unpack all parts, inspect for missing parts by packing list.

4. Set-up

The LABOROTA 20 control with its

transportation rods is bolted to the pallet.
Remove these screws and carry the unit to its
place of installation (CAUTION: 2 persons
needed; in this configuration, unit’s weight is
about 110 kg).

Place the LR 20 on a stable level stand. In

ready-to-

go configuration, unit’s weight will be

about 160 kg ! Ideal stand height above
ground is 60 cm, this enables easy access to
all controls. We recommend the optional
LABOROTA 20 base (P/N 591-23000-00).
Remove transportation rods and keep for later
use. Re-install LABOROTA 20 S control
collecting flask panels (1 & 2) having been
removed for transport. Attach panel, to include
front door (2) with screws at points (4).

Through-holes for carrying rods are plugged

with caps (5).

Install protective cabinet (6) on table (7), door

pointing towards operator. Use screws (8)
pre-installed in attaching holes.

1

6

7

4

2

4

5

1

3

8