Depress Select and Start Stop keys near Display I together and hold for 5 seconds, turn master

switch ON, release keys

„CAL“ is shown in the Display Speed, „VAC“ in the Display Time

750 is shown in Display I and 1 is shown in Display II (1st step).

Now, set the pressure measured with the separate vacuum

gauge in Display I, by pressing the Set increase key or Set
decrease key

Confirm with Start Stop key located next to Display I

Display I to show 50 and display II 2 (2

nd

step).

At this point, the LABOROTA 20 control automatically

opens the vacuum valve and evacuates until reaching
vacuum limit (the value on the separate vacuum gauge
won’t change anymore)

In Display I, set the pressure measured with the separate

vacuum gauge in Display I, by pressing the Set increase
key or Set decrease key

Confirm with Start Stop key located next to Display I. As

confirmation for the accepted calibration values, „0“
appears in Display I and Display II

Turn master switch (Power) OFF and ON

Calibration is complete